Alarm Range- 30-400fpm(15/20 m/s)
Accuracy – face velocity accuracy +/- 10%
Airflow sensor – on board or remote sensor
Calibration -Single or 2-point (installer selectable)
Low air Alar delay -fixed 5 sec
Relay output -1-(low air alarm)
Analog output – not available
Relay input -2- nigth setback and sash high
Comm.port – RS232- can be connected via serial interface to LAN network( full load performance software reporting availabiliy)
Sash high indecator
Yes Using a Micro switch aproximity switch input with repeat alarm feature factory set to 5 min. ( Can be ajustedvia laptop up to 30mins.)
Night set back Yes using a relay input
External alarm indication Not available
Power Requirement Input-120vac, 60hz
output- 15vdc,500ma
Display -visual LED: red.alarm; green,normal
Alarm Indication Red LED and Audible alarm
Home silence Yes-Temporarily
Mouthing Semi Flush
Operating Temperature 55 – 86 F (13- 30 c)
Storage Temperature -40-150 F (40-65 C)
Instrument Di mentions Instrument case: 5.2″H x 3.19″ W x 1.97″ D
(132mm H x 81mm W x 50mm D)
Agency Listings UL and CE
Display Range 0-999 fpm(0-5.0 m/s)
Alarm range 0-999fpm(0-5.0 m/s)
Field Set-up 2 -points velocity calibration (with on- screen instruction)
Acurracy Sensor/ display resolution 1 fpm
Face velocity Accuracy +/-10%
Alarm Delays User Configurable-0 to 60 secs
Relay output 4(1 on board- 3 in optional plug in relay interface unit)
Analog output 0-10v output, directly proportion to velocity (optional)
Relay input input 3
Comm.Port RS232 can be connected via serial interface to LAN network
(full software available)
Sash high indication yes with separate plug in connected
Night setback Yes
External Alarm Indication Yes
Power requirement Input 120vac,60hz
Output 15vdc,500ma
Units English and Metric (User selectable)
Display Visual Analog Bar graph of fault timeline
LED: red.alarm;Yellow, caution; green,normal
Digital Display of velocity reading (can be turn off)
Alarm Indication Red LED and Audible alarm (to a certain sound spec)
Home silence Yes-(Temporarily/permanent/automatic depending on the type of alarm or external input)
Mouthing Semi Flush
Operating Temperature 55 – 86 F (13- 30 c)
Storage Temperature -40-150 F (40-65 C)
Instrument Di mentions Instrument case: 5.2″H x 3.19″ W x 1.14″ D
(132mm H x 81mm W x 29mm D)
Agency Listings UL and CE